1 · Why this matters

Established The prize is knowing which of the more than 1,600 AI-enabled devices authorised in the United States change what happens to patients, by how much, for whom and for how long. Today that is known for a handful. Where it has been measured properly the effect is specific and worth having: in a randomised screening trial of 105,934 women, AI-supported reading raised sensitivity from 73.8 to 80.5 per cent at an unchanged specificity of 98.5 per cent while cutting radiologists' screen-reading workload by 44 per cent, and in a 2026 meta-analysis of cardiovascular trials, image-based decision support carried a risk ratio of 0.74 for major adverse cardiovascular events.

Established The baseline is authorisation without measurement, and its cost is already visible in the record. In a 903-device sample, 55.9 per cent of devices had a clinical performance study at clearance; in a 692-device review, 3.6 per cent of approvals reported the race or ethnicity of the people a device was tested on; 43.4 per cent of recalls of cleared AI devices came within twelve months of clearance, about double the rate for 510(k) devices generally. Frontier A validation regime earns its cost in the failures it finds before patients do. The clearest case is a popular proprietary sepsis model that scored an area under the curve of 0.87 when every prediction up to discharge was counted and 0.47 once predictions made after clinicians had already begun treatment were excluded: a tool that mostly repeated what the ward already knew, found by an independent team rather than by any regulator.

Established The gain from validation does not fall evenly, and the evidence says who is missing from it. Of 86 randomised trials of clinical AI counted to late 2023, 22 reported race or ethnicity at all, with a median of 70.5 per cent White participants. Canada's April 2026 premarket guidance asks manufacturers to show performance for subgroups, including at their intersections, which is a stated expectation rather than yet a measured result. Speculative Large health systems with informatics staff can test a model silently before switching it on; the smaller hospitals that buy the same products mostly inherit whatever the vendor's evidence says.

Frontier The condition is a measurement system that no jurisdiction yet operates. The upside exists only if someone records which model version scored which patient, and what happened to that patient afterwards, at a scale where drift and subgroup failure become visible. No regime this brief examined requires that record in November 2026; the EU's logging obligation for high-risk AI would be the nearest, and for medical devices it is not due before August 2028. Section 5 explains why the record does not exist by default.

2 · Concept overview

Established Clinical AI validation is the chain of evidence connecting a model's output to what happens to a patient, and it has four links, each separately measurable. Technical validity: does the model reproduce its reference labels on data it was not trained on. Clinical validity at the point of use: does it perform on this hospital's population, scanners, coding habits and case mix. Clinical utility: when clinicians act on it, do patients do better than when they do not. Durability: does it still do so a year later, after the population, the record system and the model itself have changed. Frontier A regulatory authorisation speaks mainly to the first link, partly to the second, rarely to the third and, in the regimes this brief examines, almost never to the fourth.

Established The United States list is the largest public count of authorised clinical AI, and it is a count of regulatory decisions. The FDA states that it had authorised over 1,600 AI-enabled devices for marketing as of September 2026. Devices are placed on the list primarily because AI-related terms appear in the summary descriptions of their authorisation documents, and the agency says the list is not a comprehensive resource. Established Most of those decisions are 510(k) clearances, a pathway that asks whether a device is substantially equivalent to one already legally marketed, not whether patients do better with it. Frontier The count therefore measures throughput. It carries no information about effect on patients until it is joined to an outcome record, and section 3 shows that the record is mostly absent.

Frontier Three kinds of system fail validation in three different ways, and this brief keeps them apart. Narrow imaging and signal devices, 76.6 per cent of them in radiology in one 903-device sample, have fixed inputs and labels that can be checked against biopsy or follow-up; their characteristic failure is a shift in the population or the equipment. Risk scores built on the electronic health record ingest the very workflow they are meant to inform; their characteristic failure is learning the clinicians' own suspicion and reporting it back. General-purpose language models used by clinicians and the public act through conversation; their characteristic failure is that a model which knows the answer and a person using it together reach the wrong one. Established Each failure has now been measured at least once, and section 3 gives the instance.

Established Neighbouring briefs own adjacent pieces, and this one does not restate them. Precision Medicine reads the Swedish MASAI screening trial's detection and workload figures as workflow evidence; Human-AI Integration holds the German PRAIM programme as the strongest observational case of human-machine complementarity; Intelligence Measurement owns the validity of benchmarks in general; AI Governance owns the EU AI Act's calendar and its evaluators. Frontier None owns the question here: what an authorisation of a clinical model is evidence of, how long that evidence stays true once the model is deployed, whom it fails, and who finds out when it does.

3 · Current scientific position

Established The count is large, recent and administrative. The FDA states that it had authorised over 1,600 AI-enabled medical devices for marketing in the United States as of September 2026. A Canadian law firm's April 2025 account put the figure at 1,016 as of the agency's update of 25 March 2025, so on this brief's arithmetic roughly 600 were added in about eighteen months, some 33 a month. Established The list is assembled from AI-related terms in the summary descriptions of authorisation documents, considered alongside device classification, and the agency states that it is not a comprehensive resource. Established The FDA says it will explore methods to identify and tag devices that incorporate foundation models, and encourages sponsors to say so in their public summaries; it has not yet done so, which means the public list cannot tell a reader which authorised devices contain a large language model.

Established Almost all of it arrived by resemblance rather than by trial. Siontis and colleagues at the University of Bern examined 903 FDA-cleared AI-enabled devices, 76.6 per cent of them used in radiology, in JAMA Network Open in April 2025. On the trade-press account of the paper, 877 were cleared through the 510(k) pathway and 22 by De Novo classification: about 97 per cent by substantial equivalence, on this brief's arithmetic. Established Clinical performance studies were available for 505 devices, 55.9 per cent, at the time of clearance, and retrospective designs were the most common among those that reported one. Established Information on sex subgroups was available in 145 of those 505 studies, 28.7 per cent, and on age subgroups in 117, 23.2 per cent; these are shares of the devices that had a study at all, not of the 903. Frontier The authors' own reading is that substantial equivalence may be an inadequate marker of safety and efficacy. This brief's reading is narrower: it is a marker of something else, resemblance to a predecessor whose own evidence may have been equally thin.

Established What the public summaries disclose is thinner still. Muralidharan and colleagues reviewed 692 FDA-approved AI devices authorised between 1995 and 2023. On their abstract, 3.6 per cent of approvals reported race or ethnicity, 99.1 per cent gave no socioeconomic data, 81.6 per cent did not report the age of study subjects, 46.1 per cent gave comprehensive detailed results of performance studies, 1.9 per cent linked to a scientific publication with safety and efficacy data, and 9.0 per cent contained a prospective study for post-market surveillance. Frontier An absence of demographic data is not evidence of bias. It is the absence of the data that would show whether there is any, and it is the default condition of the public record.

Established Recalls arrive early, and they cluster where clinical validation is missing. Lee, Kramer, Sandri and colleagues, with Ross, Sharfstein and Dai, matched 950 cleared AI devices against recall records in November 2024. Sixty devices, 6.3 per cent, accounted for 182 recall events, and 79 of those recalls, 43.4 per cent, came within the first twelve months after clearance, approximately double the rate reported for all 510(k) devices. Established The manufacturer-attributed causes were diagnostic or measurement errors in 109 recalls, functionality delay or loss in 44, physical hazards in 14 and biochemical hazards in 13. Established Among recalled devices, 40 per cent of those from privately held companies lacked clinical validation, against 77.7 per cent from established public companies and 96.9 per cent from smaller public companies, and devices without reported validation had more recalls per device, a mean of 3.4. Public companies made 53.2 per cent of the devices and accounted for 91.8 per cent of recalls and 98.7 per cent of recalled units. Frontier A recall is a lagging, manufacturer-reported signal of a failure somebody noticed. A model that degrades slowly, or fails only for a subgroup, may produce no recall at all, so 6.3 per cent is a floor on failure rather than an estimate of it.

Established The randomised record has been counted three times, and each count is small. Lam, Cheung, Munro, Lim, Shung and Sung screened 11,839 articles in 2022 and found 39 randomised trials of AI in clinical practice, 0.33 per cent of what their search returned; the AI arm outperformed usual care in 30, and clinically relevant outcomes improved in 21 of those 30, with gastroenterology, including 15 trials of AI-assisted endoscopy, the largest group. Established Han, Acosta, Shakeri, Ioannidis, Topol and Rajpurkar counted 86 unique trials between January 2018 and November 2023. Seventy reported favourable primary endpoints; 92 per cent were run in a single country, primarily the United States or China; gastroenterology made up 43 per cent, radiology 13, surgery 6 and cardiology 5; and the primary endpoint was diagnostic performance or yield in 46 trials, care management in 18, patient symptoms and behaviour in 15 and clinical decision-making in 7. Established Race or ethnicity was reported in 22 of the 86, with a median of 70.5 per cent White participants. Frontier On this brief's arithmetic the list grew by about 33 devices a month in 2025 and 2026, while Han's count implies about 1.2 randomised trials a month from 2018 to 2023. The periods differ and a trial of an AI intervention is not necessarily a trial of an authorised device; beyond the review in the next paragraph, which found regulatory approvals for six of the systems it studied, this brief found no count of how far the two populations overlap.

Established When benefit and harm are counted together, harm is mostly not counted. A systematic review in The Lancet Regional Health – Europe of AI-related algorithmic decision-making systems used by healthcare professionals, as News-Medical reported it in December 2024, included 19 studies, 18 randomised trials and one prospective cohort. Twelve reported patient-relevant benefits; only eight included standardised harm assessments; risk of bias was low in four trials, moderate in seven and high in seven. Established Six of the systems had regulatory approvals, and the associations between regulatory status, study quality and patient outcomes remained inconclusive. Frontier That last finding is the nearest thing in the literature this brief read to a direct test of whether authorisation predicts benefit. It returned no answer, which is the honest state of the question.

Established The best hard-outcome signal in a large clinical area is one pooled estimate at low certainty. Ong, Johnson and Leeson's systematic review and meta-analysis of randomised trials of AI in cardiovascular care, dated August 2026, included 31 trials from 13 regions with 1,685,717 patients, 11 of them, with 1,614,689 patients, in the meta-analysis. Ninety per cent were published after 2021, 58 per cent were multicentre, and 65 per cent followed patients for under twelve months. Established Twenty-two trials, 71 per cent, reported a significant benefit on their primary endpoint, mostly on intermediate process measures, and risk of bias was low in seven, 23 per cent. Established The one hard clinical outcome pooled was major adverse cardiovascular events, for which image-based AI decision support carried a risk ratio of 0.74 (95 per cent CI 0.58 to 0.96; I² 0 per cent). AI-enhanced electrocardiograms had no significant effect on heart-failure detection (risk ratio 1.22, 0.70 to 2.14; I² 90 per cent), and certainty of evidence ranged from low to very low. Frontier A 26 per cent relative reduction in major cardiac events is a large number if it holds. It rests on a subset of trials whose reviewers themselves call for longer, more rigorous and more transparently reported follow-up.

Established The strongest randomised result in the field measured the patient endpoint, and it came back non-inferior. The final MASAI analysis, by Gommers, Hernström and colleagues with Lång as senior author, published in The Lancet in January 2026, randomised 105,934 women to AI-supported screening or to standard double reading without AI. The interval cancer rate was 1.55 per 1,000 participants with AI (95 per cent CI 1.23 to 1.92) against 1.76 per 1,000 without (1.42 to 2.15), a proportion ratio of 0.88 (0.65 to 1.18; p = 0.41), non-inferior against a pre-specified 20 per cent margin. Established Sensitivity was 80.5 per cent with AI against 73.8 per cent (p = 0.031), the advantage holding across age and breast-density groups and for invasive but not in-situ cancers; specificity was 98.5 per cent in both arms. Established Lund University's release reported 82 interval cancers against 93, described as 12 per cent fewer, with 16 per cent fewer invasive, 19 per cent fewer large and 27 per cent fewer aggressive-subtype interval cancers, and carried forward from earlier reports a 29 per cent increase in detected cancers and a 44 per cent reduction in screen-reading workload (trial team's institution). Frontier The 12 per cent is the point estimate of a ratio whose interval runs from a 35 per cent reduction to an 18 per cent increase. The trial was designed to show that AI-supported reading is not worse, and it showed that; it did not show that it is better on the endpoint patients experience. Frontier What this brief adds to its neighbours' coverage is only this: the patient-relevant endpoint has now been measured at this scale in one randomised trial, in one screening programme.

Established A popular hospital sepsis model was measured independently twice, and the second measurement explains the first. In the first, a Michigan Medicine team including Wong and Singh examined close to 40,000 hospitalisations from 2018 and 2019 and reported in JAMA Internal Medicine an area under the curve of 0.63, against the 0.76 to 0.83 reported by the developer, Epic Systems. On the university health-policy institute's account of the study, the model alerted on nearly one patient in five, the chance of sepsis after an alert was 12 per cent, and eight patients had to be evaluated to find one case. Established In the second, Kamran, Tjandra and colleagues at the same institution, in NEJM AI in February 2024, evaluated 77,582 adult hospitalisations from 2018 to 2020, 3,766 of them, 4.9 per cent, with sepsis. Counting every prediction up to discharge the model scored an AUROC of 0.87; restricted to predictions before sepsis criteria were met it scored 0.62; and when predictions after clinical recognition were excluded it scored 0.47 (95 per cent CI 0.46 to 0.48). Established In 3,193 of the sepsis hospitalisations, 84.8 per cent, a treatment indicator such as a blood-culture or fluid order came before the sepsis criteria were met, and the authors attribute most of the model's discriminative ability in standard evaluation to predictions made after clinicians had recognised sepsis and begun treatment. Frontier This is the signature failure of a model trained on the record of care: it learns the clinicians' suspicion from their orders, scores well against outcomes the clinicians had already acted on, and is least informative at the one moment it was bought for. An evaluation that does not fix the time of prediction relative to the clinical action measures the hospital rather than the model.

Established Drift has been measured across a decade of data, and its largest single cause in that measurement was a change in diagnosis codes. Yang, Karstens, Ross and Yala rebuilt approximations of two commercial sepsis systems from their published feature lists on the MIMIC-IV intensive-care dataset and tracked performance over ten years: a recurrent network on the Epic feature set fell from an AUC of 0.729 to 0.525, a logistic regression on the same features from 0.706 to 0.626, and models on the Dascena feature set declined much less, from 0.723 to 0.707 and from 0.687 to 0.640. Established The authors identify the 2015 switch from ICD-9 to ICD-10 diagnosis codes as a major cause; retraining without the code-based features reduced the drop by 0.15. They also record practice drift beneath it: average sepsis onset time rose from 9 to 12 hours by the final period, and early microbiology sampling fell, MRSA screens by 86 per cent. Frontier These are replications built from feature lists, not the commercial models, and they omit any retraining a hospital might do. The mechanism they expose, an administrative change silently breaking a clinical model, is still one that no premarket test could have seen, because at the time of the test it had not happened.

Established A randomised, preregistered test of language models with real users found that people using a model which knew the answer did no better than people without one. Bean, Payne, Mahdi and colleagues at Oxford randomised 1,298 members of the public across ten medical scenarios to GPT-4o, Llama 3, Command R+ or a control condition using sources of their own choosing, reporting in Nature Medicine in 2026. Established Tested alone, the models identified the underlying conditions in 94.9 per cent of cases and the correct disposition in 56.3 per cent on average; participants using them identified relevant conditions in fewer than 34.5 per cent of cases and the correct disposition in fewer than 44.2 per cent, no better than the control group. Established The authors conclude that standard benchmarks for medical knowledge and simulated patient interactions do not predict the failures they observed with human participants. Frontier This is the general-purpose version of the sepsis result: the performance that matters belongs to the person-and-model system, it can be measured, this trial measured it, and it is not the number any benchmark or authorisation summary reports. The wider argument about benchmark validity belongs to the Intelligence Measurement brief.

Established The adverse-event system cannot say, one time in three, whether an algorithm was involved. Handley, Krevat, Fong and Ratwani reviewed 429 reports concerning AI- or machine-learning-enabled devices in the FDA's device-report database: 108, 25.2 per cent, were potentially related to the AI or machine-learning function, and 148, 34.5 per cent, lacked enough information to tell. Frontier Babic, Cohen, Stern and colleagues, reviewing the post-market surveillance record for about 950 devices authorised between 2010 and 2023, conclude that the existing system is insufficient for properly assessing the safety and effectiveness of AI and machine-learning devices. Frontier The instrument meant to detect harm after authorisation records events; it was not built to attribute a confident wrong output to a model version, and the record it produces cannot attribute the failure this brief is about.

4 · Frontier questions

Frontier Does authorisation predict local performance at all? This brief found no published study that takes a sample of authorised devices and re-measures them at independent sites against their stated performance. The nearest evidence points one way without settling anything: in the recall data, devices without reported clinical validation had more recalls per device, while the Lancet Regional Health review found the association between regulatory status and patient outcome inconclusive. Speculative If authorisation carries real information about local performance, the 1,600 count means more than this brief allows; nothing measured so far shows that it does.

Frontier Is non-inferiority the right bar for AI that substitutes for a clinician? MASAI was designed to show that replacing one of two human readers does not worsen interval cancers, and the operational case for adoption is the reading workload saved. Speculative If substitution is the business case, trials powered for superiority on patient endpoints may never be funded, because the buyer's question has already been answered. Frontier Against that, a sensitivity gain of 6.7 percentage points at equal specificity is a superiority result on an intermediate endpoint, and whether it becomes fewer deaths is a question only longer follow-up can answer.

Frontier Can drift be detected before it harms anyone? The FDA's 2025 request for comment states the problem in the regulator's own words: performance can change with shifts in clinical practice, patient demographics, data inputs and health-care infrastructure, which may lead to performance degradation, bias or reduced reliability, and retrospective testing and static benchmarks are not designed to predict behaviour in dynamic, real-world environments. Frontier It asked developers and health systems what triggers the need for more intensive evaluation and how they define and respond to degradation, questions to which there is no agreed answer. Speculative The hard part is the label. A screening model's misses surface as interval cancers only after the screening interval has run, and a sepsis model's labels are partly written by the treatment it prompts, so a drift monitor that waits for outcomes is late and one that does not wait is measuring proxies.

Frontier Is measured subgroup disparity in the model or in the labels? Seyyed-Kalantari, Zhang, McDermott, Chen and Ghassemi reported in 2021 that chest-radiograph classifiers trained on three datasets consistently yielded higher false-positive rates for the “no finding” label, which is to say more missed disease, in subgroups known to be historically underserved. Frontier Bernhardt, Jones and Glocker replied that the setup cannot establish underdiagnosis: the test data carried the same bias as the training data, the labels were extracted from reports by natural-language processing with an F1 score as low as 0.76 for “no finding” on one benchmark, and underdiagnosis is an annotation shift that re-sampling, re-weighting and threshold calibration cannot correct, so the effective remedy is re-annotation. Frontier The Institute declares a tie on cause and not on consequence: a model that inherits underdiagnosis from its labels will reproduce it in practice whichever side is right, and the experiment that separates the two readings is a re-annotated, subgroup-labelled test set, which section 7 asks for.

Frontier How is a model validated when it is designed to change? The predetermined change control plan is the regulators' answer for planned change: the modifications, the method for making and validating them, and an assessment of their impact are reviewed in advance, within the device's intended use. Speculative It has no settled answer for change the manufacturer does not control. DLA Piper's December 2025 analysis lists, as matters still requiring guidance, validation when core AI components are hosted externally and the risks of continuously learning models and weight updates; a device built on a third-party foundation model inherits that model's update schedule.

Frontier Does local validation scale? Youssef, Pencina, Thakur, Zhu, Clifton and Shah argue that external validation “neither guarantees generalizability nor equates to a model's clinical usefulness”, and propose recurring local validation: site-specific reliability tests before every deployment, then regular and recurrent checks, with a silent mode in which outputs are recorded and compared with the clinical ground truth before anyone acts on them. Speculative Every hospital revalidating every model is the correct procedure and an expensive one, its cost falls hardest on the institutions least able to pay it, and whether a shared or third-party validation market could carry it is unmeasured.

Frontier What does AI assistance do to the clinician when the AI is wrong? The Intelligence Amplification brief records the settled taxonomy of automation bias, a cited breast-cancer study in which cancers found in 46 per cent of cases without an automated aid were found in only 21 per cent when the aid missed them, and a contested first measurement of AI-induced deskilling. Frontier No authorisation pathway examined here requires a measurement of the person-and-model error rate in the cases where the model errs, which is the number that decides whether a wrong output is caught.

5 · Technological bottlenecks

Established There is no denominator. The FDA list counts authorisation decisions. It does not count installations, model versions, sites or patients scored, and nothing in the material this brief examined does. Frontier Without a deployment record no failure rate can be computed for any device, and the 182 recall events in the Lee dataset cannot be expressed per patient, per site or per year of use.

Established The labels are written by the care the model is meant to change. In the Kamran evaluation, a treatment indicator preceded the sepsis criteria in 84.8 per cent of sepsis hospitalisations. Frontier For any model trained on the electronic record, the outcome label partly encodes the clinicians' response to the patient. Speculative Once such a model is deployed and changes practice, the label also begins to encode the model's own influence, so a successful model contaminates the data its successor will be trained and validated on.

Established Patient endpoints arrive late. An interval cancer is by definition one that surfaces between screening rounds, so the endpoint that judges a reading arrives only after the interval has run; MASAI's final interval-cancer analysis was published in January 2026, after the trial's detection and workload results had already been reported. Frontier For every AI device whose benefit is a downstream outcome, validation of utility lags deployment by the natural history of the outcome.

Established The incident system cannot attribute algorithmic failure. One report in three in the Handley sample lacked the information to say whether the AI function contributed, and Babic and colleagues judge the system insufficient for AI devices. Speculative A report that does not record which model version produced which output on which input leaves an algorithmic failure indistinguishable from user error, and the clinician who acted on the output is the person most likely to be named.

Established The subgroup data are missing at source. Race or ethnicity appears in 3.6 per cent of the approvals Muralidharan's team reviewed and in 22 of the 86 trials Han's team counted; sex and age subgroup results appear for under a third and under a quarter of the devices with any clinical study in the Siontis sample. Frontier A subgroup failure cannot be detected after deployment in data that never recorded the subgroup.

Frontier Proprietary models resist the independent measurement that found the problem. Singh's conclusion from the Michigan validation was the need for more open-source models that can be easily externally validated. Frontier The vendor's response, as Youssef and colleagues record it, was to announce that a new version would be localised to each site's patient mix. That is the right engineering, and it makes every installation a different model: the published evaluation of the old version describes none of them.

Speculative The evidence costs fall on the party that does not need the evidence to sell. A device cleared by substantial equivalence can be sold from the day of clearance, while the randomised or post-market evidence that would establish its effect is a cost to whoever chooses to generate it. Frontier The recall data show the pattern without explaining it: public companies supplied 53.2 per cent of the devices and 91.8 per cent of the recalls.

6 · Research dependencies

Established The subject depends on measurement science it does not itself produce. Confidence intervals, contamination control and named reference populations for evaluation results are the ground the Intelligence Measurement brief covers. Frontier Clinical validation inherits every one of those problems and adds one: its labels are clinical outcomes, which carry their own error and are partly produced by the care being evaluated.

Frontier It depends on causal methods for models that alter the outcomes they predict. The correction in the Kamran evaluation, scoring only predictions made before treatment indicators, is a design choice rather than a method anyone has standardised. A general method for evaluating a deployed model whose outputs change clinicians' actions, and therefore the labels, is an open research question.

Frontier It depends on label quality that no regulator specifies. The Bernhardt critique turns on label noise, an F1 score as low as 0.76 for one label, and argues that only re-annotation corrects annotation shift. Speculative Re-annotating at scale requires expert reader time that health systems do not budget for validation, which is why the remedy is known and rarely applied.

Frontier In Europe it depends on standards that are not yet written. The AI Governance brief records that the harmonised standards on logging, human oversight and accuracy were still in preparation in October 2026 under a request running to 28 February 2027, against high-risk obligations deferred to 2 August 2028 for Annex I products. Frontier Without those standards the logging and oversight obligations that would apply to AI medical devices have nothing an engineer can build against or an assessor can audit against.

Frontier It depends on human-factors evidence from outside medicine. The automation-bias literature reviewed in the Intelligence Amplification brief predates clinical AI and supplies the vocabulary of commission and omission errors. What clinical validation needs from it is a standard way to measure the person-and-model error rate in the cases where the model is wrong.

7 · Required experiments

Frontier The decisive experiment is a random-sample re-validation of authorised devices at sites that did not build them. Draw devices at random from the FDA list, stratified by review panel, and run each one silently at several independent health systems against outcome labels fixed before deployment. Then report, for each device, whether it meets the performance its authorisation summary states, overall and by sex, age and race, at deployment and again a year later. Frontier If most devices meet their stated performance and hold it across subgroups and over the year, this brief's central claim narrows from “authorisation says little about effect” to “authorisation says little about utility”, and the case for mandatory post-market performance reporting weakens accordingly; if a material share do not, that case stops being an argument and becomes a measured requirement. Established Nothing in the material this brief examined schedules or funds such a study, and the FDA's 2025 request for comment on real-world performance asked how such measurement should be done rather than committing to do it, so the experiment is unscheduled. Established It needs no new technology: silent-mode operation is already described in the local-validation literature, and each participating site needs only the device, its own records and a fixed labelling protocol.

Frontier A second patient-endpoint trial, outside the first trial's setting. MASAI showed that a randomised trial of more than 100,000 people with an interval-cancer endpoint can be run inside a population screening programme. Speculative Whether its result generalises is a question only a second trial in another country's programme can answer. Without one, non-inferiority in a single programme is the only randomised interval-cancer result this brief has read, and it should be quoted as exactly that.

Frontier Time-locked evaluation of every electronic-record model already in use. The Kamran design, scoring only predictions made before the first treatment indicator, can be applied by any hospital that runs a model and keeps its own records. Established It needs no new data and no trial, and in the published application this brief read it moved the measured AUROC of a widely used model from 0.87 to 0.47. Speculative This brief found no systematic application of the design across vendors' models, and expects it to move others.

Frontier A drift-detection test against a known shift. The ICD-9 to ICD-10 transition is a natural experiment with a known date, and AI Gone Astray reconstructs, after the fact, AUC drops of 0.08 to 0.20 for models on the Epic feature set. A monitoring method can be scored on how quickly it would have flagged that degradation from live data. Speculative A method that cannot detect a shift of that size within months of an announced coding change cannot be relied on to detect the shifts nobody announces.

Frontier Human-user testing before public release of any language-model health tool. The Bean trial is the template: a randomised comparison of users with and without the model on fixed scenarios with known answers. Established Its authors recommend systematic testing with human users before deployment in public healthcare. Speculative Requiring it would cost weeks per model; not requiring it means the 94.9 per cent figure is the one that gets quoted.

Frontier Error-injection reader studies to measure automation bias. Present clinicians with cases in which the AI output has been made wrong at a known rate, and measure the person-and-model error rate on those cases against unaided reading. Speculative The design is standard in human-factors work and cheap to run, and it would give every authorisation a second number, how often a wrong output is caught, that no summary now reports.

Frontier A re-annotated, subgroup-labelled test set for the underdiagnosis question. Expert re-reading of a stratified sample of the radiographs at issue, blinded to subgroup, would separate algorithmic bias from label bias, the remedy Bernhardt, Jones and Glocker propose. Speculative The result would settle, in either direction, a tie the Institute currently declares.

Established A natural experiment in regulation is already running. Health Canada's guidance, published on 1 April 2026, asks manufacturers of Class II to IV machine-learning devices for evidence of performance for appropriate subgroups, including at the relevant intersections, and for data adequately representative of the Canadian population and clinical practice. The transparency principles the same agency issued with the FDA and the MHRA list known biases and failure modes as good practice to communicate, and are not binding. Speculative Whether the Canadian expectation produces subgroup evidence that the same devices' US summaries lack is a comparison that becomes possible as Canadian licence decisions under the guidance accumulate, and it is the cheapest available test of whether a premarket subgroup requirement changes what is known.

8 · Engineering requirements

Established The change control plan is the one piece of lifecycle engineering the regulators have specified. The FDA's final guidance on predetermined change control plans for AI-enabled device software functions, in the version dated August 2025, says a plan should describe the planned modifications, the methodology to develop, validate and implement them, and an assessment of their impact, and it applies across the 510(k), De Novo and premarket approval pathways. Established Health Canada's April 2026 guidance uses the same three parts, a change description, a change protocol and an impact assessment. Changes made under an authorised plan need no significant-change amendment, while changes to the device's medical conditions, purposes or uses require a licence amendment before they are implemented. Established The five principles the FDA, Health Canada and the MHRA identified jointly, focused and bounded, risk-based, evidence-based, transparent, and a total-product-lifecycle perspective, limit planned changes to modifications within the device's intended use.

Frontier A change control plan is not continuous learning in the field. Every change it permits is specified, bounded and validated by a protocol written before authorisation, so the model that runs after an update is a pre-approved variant rather than a system that learned from last week's patients. Established The FDA's lifecycle-management guidance, which would set expectations for the rest of the device's life, was still marked as a draft dated January 2025 on the agency's page in late 2026.

Frontier The monitoring stack that validation requires has five parts, and none is exotic. A version-stamped log of every output against the patient and input it was produced for; a labelling pipeline with a fixed delay and a fixed definition, set before deployment; stratification of performance by the subgroups the training data may under-represent; pre-registered triggers for re-evaluation and rollback; and a reporting channel that can attribute a failure to a model version. Established The FDA's 2025 request for comment asks about nearly each of these in turn, the metrics used, the monitoring tools, the balance of human and automated review, the triggers for more intensive evaluation and the response to degradation, which is a fair description of what does not yet exist as a requirement. Speculative The engineering is ordinary software practice; what is missing is an obligation, and someone to pay for the labels. The Precision Medicine brief names the same missing obligation as a constraint on its own subject.

Frontier Silent mode is the cheapest validation tool there is. Running a model in the background, recording its outputs and comparing them with what happened before anyone acts on them, is the procedure Youssef and colleagues place inside recurring local validation. Speculative It costs compute and an analyst's time; it cannot measure utility, because nobody acts on the output; and it is the only way to measure clinical validity at a site without exposing patients to an unvalidated model.

Frontier Localisation trades generalisability for local performance, and moves the validation burden to the site. A model tuned to each hospital's patient mix should perform better there, and each tuned copy needs its own validation. Speculative For devices built on third-party foundation models the problem compounds: the version that was validated may not be the version that is running unless the manufacturer pins it, and the regimes examined here do not yet say how pinning is to be demonstrated. The human side of the same deployment, where the clearest evidence of complementarity sits, is the subject of the Human-AI Integration brief.

9 · Adjacent technologies

Established Intelligence Measurement is the general theory of which this brief is a clinical case. Its argument that an evaluation is an experiment, reported without intervals or named reference populations, applies to every authorisation summary discussed here; the Bean trial is the clinical instance of its benchmark-to-use gap.

Established Human-AI Integration and Intelligence Amplification own the human half of the system. The first holds the German PRAIM screening programme as the best observational evidence that clinician and model together can raise detection and lower recall at once; the second holds automation bias and deskilling. Frontier Clinical validation needs both, because the performance that reaches a patient belongs to the clinician-and-model pair and not to the model.

Established Precision Medicine faces the same structure one layer down. A companion diagnostic that is right about the target and wrong about the patient is the pharmacological version of an accurate model that does not change outcomes, and that brief's reading of MASAI as labour substitution rather than new diagnostic capability is consistent with the interval-cancer result recorded here.

Established AI Governance owns the EU AI Act and the public evaluators. Its finding that the general-purpose evaluators hold limited, voluntary access is the frontier-model version of this brief's finding about proprietary clinical models. Frontier In both, the party best placed to measure performance is the one that sells the system.

10 · Institutional requirements

Established United States: three premarket pathways, one of which carries nearly all the volume, and a post-market regime still at the request-for-comment stage. Devices reach the US market by 510(k) clearance, De Novo classification or premarket approval; in the 903-device sample, 877 were 510(k) clearances and 22 De Novo. Established The agency's AI-specific instruments are the change control plan guidance, final in its August 2025 version; the lifecycle-management guidance, a draft since January 2025; a request for public comment on measuring and evaluating real-world performance, docket FDA-2025-N-4203, whose comment period closed on 1 December 2025; and a discussion paper on generative-AI-enabled devices, docket FDA-2026-N-7874, whose comments were due on 19 October 2026. Established Both of the last two describe themselves as for discussion only, not draft or final guidance, and not intended to propose or implement policy changes. Established The Digital Health Advisory Committee met on the total product lifecycle of generative-AI-enabled devices on 20 and 21 November 2024, and on generative-AI-enabled digital mental health devices on 6 November 2025. Frontier The pattern is a regulator that has specified, in guidance, how a manufacturer may change a model, and has asked, in a docket, how anyone would know whether the model still works.

Established Canada: a final guidance with the most explicit subgroup expectation in any text this brief examined, and no federal AI statute. Health Canada's pre-market guidance for machine-learning-enabled medical devices, published on 1 April 2026, applies to new and amendment applications for Class II, III and IV devices. It asks for data adequately representative of the Canadian population and clinical practice, for representation across sex, gender, race, age and other identity factors, for sex- and gender-based analysis where appropriate, and for evidence of performance for appropriate subgroups, including at the relevant intersections. Established It names performance degradation over time and shifts in input data after deployment as concerns and asks manufacturers to consider describing performance monitoring plans; the Minister may impose or amend terms and conditions on a licence under subsection 36(2) of the Medical Devices Regulations, and the guidance gives machine learning and change control plans as an example of the uncertainty such terms can manage. Established On a Canadian law firm's account, the federal Artificial Intelligence and Data Act died on the Order Paper when Parliament was prorogued on 6 January 2025, and Ontario's Enhancing Digital Security and Trust Act, 2024, which could reach hospitals as prescribed public-sector entities, had royal assent and no coming-into-force date as of April 2025. Frontier “Consider describing” a monitoring plan is a premarket expectation, not a post-market reporting obligation, and the Canadian regime, like the American one, measures what is submitted rather than what happens afterwards.

Established European Union: the AI Act was to add a second layer on top of the device regulations, and both the layer and its date are moving. Medical-device AI is classed as high-risk under Article 6(1) of the AI Act through Annex I, whose Section A lists the Medical Device and In Vitro Diagnostic Regulations at points 11 and 12. The Medical Device Coordination Group's June 2025 guidance, MDCG 2025-6, says such systems must comply with both regimes, that manufacturers can integrate AI Act obligations into existing device processes, and that it introduces no obligations beyond those already in the texts. Established The Commission's Digital Omnibus proposal of 19 November 2025 would have applied the Annex I obligations from the earlier of twelve months after a decision confirming that standards and guidance exist, or 2 August 2028; MedTech Europe argued for 2 August 2029 (industry association); and the AI Governance brief records that the adopted text fixed 2 August 2028. Established Separately, the Commission's December 2025 proposal to simplify the device regulations, file 2025/0404(COD), would delete the two device regulations from Section A of the AI Act's Annex I and add them to Section B, the move applying on the new regulation's entry into force. In the Commission's presentation to the Council working party on 3 June 2026, the Commission would instead take the AI Act's high-risk requirements into account when adopting acts under the device regulations, AI-specific risks including bias, robustness and drift would be assessed within the device framework as part of clinical evidence and the general safety and performance requirements, and device authorities would cooperate with AI market-surveillance authorities, whose mandate covers deployers such as healthcare providers. Frontier If the amendment is adopted before August 2028, the AI Act's high-risk requirements would reach medical devices only through the device regulations, and the date the market has planned against would matter less than the content of the acts adopted under them; this brief has no record of the proposal's adoption as of November 2026.

Frontier Liability is unallocated, and this brief found no measured record of how it has been allocated in practice. The Canadian analysis cited above notes that liability could be hard to apportion among clinicians, hospitals and software manufacturers when a faulty AI output is relied on. Speculative Until it is allocated, the party best placed to validate locally, the hospital, has a weak incentive to discover a failure it might then be held to have known about, and the incentive structure favours not looking.

11 · Ethical & societal considerations

Frontier The harm validation would catch falls on the people the evidence omits. With race or ethnicity reported in 3.6 per cent of approvals and in 22 of 86 trials, and with a contested but unrefuted report of more missed disease in historically underserved subgroups on chest radiographs, the patients about whom a device's evidence says least are also those whose errors are least likely to be noticed.

Frontier Silent-mode validation raises a consent question. Running a model on patients' records without acting on it exposes them to no clinical risk, and uses their data to evaluate a commercial product. Speculative Whether such evaluation is quality improvement or research, and whether patients should be told, is a question this brief found no regulator answering.

Frontier Automation bias moves responsibility without moving accountability. If a clinician who follows a wrong output is less likely to catch it, as the Intelligence Amplification brief's figures indicate, then a device's error becomes the clinician's error in the record, and an incident system that cannot attribute algorithmic failure one time in three will file it there.

Frontier Language models used directly by the public move the validation burden onto the patient. In the Bean trial the participant was the person in the loop, and the person-and-model system did no better than the control group. Speculative A tool that reaches the public without passing through a clinician puts the user in the role the evidence shows people do not yet perform well, and the device regimes examined here govern it only if it is a device.

Speculative Equity cuts both ways, and the evidence does not say which way it will cut. AI-supported screening that cuts reading workload by 44 per cent could extend screening capacity where radiologists are scarce. It could equally be deployed first where it saves the most money, which need not be the same place.

12 · Civilizational implications

Frontier Medicine is choosing between two evidence cultures for software. Drugs reach patients through trials of the outcome; devices mostly through resemblance to a predecessor; clinical AI has arrived, at some 33 authorisations a month, through the second door. Speculative If the device culture prevails, the authorisation count will keep outgrowing the outcome record, and the working assumption that “authorised” means “shown to help” will be wrong for most of what is authorised.

Speculative If the validation infrastructure is built, the record becomes an asset nobody has had. A version-stamped record of model outputs and patient outcomes across health systems would be the first longitudinal account of how clinical decisions change when software takes part in them, and it would make each new model testable against the last. Handwave Claims that such a record would let medicine learn continuously from every patient assume that the labelling problem in section 5 is solved, which nothing yet shows.

Speculative The workforce effect is the most likely large consequence and the least measured. MASAI's 44 per cent workload reduction is a labour result. If it generalises, its effect on how many radiologists a screening programme needs will arrive long before any mortality result, and deskilling, which the Intelligence Amplification brief records as contested, will decide whether the remaining readers can catch what the model misses.

13 · Timelines

These horizons track when the gap between authorisation and measured effect could narrow, not when clinical AI will be adopted.

  • 10 yr: Established The EU AI Act's high-risk requirements are due to apply to Annex I products, which include AI medical devices, on 2 August 2028, unless the proposed device-regulation amendment moves the device regulations to Section B of Annex I first; the harmonised standards on logging, human oversight and accuracy are due under a request running to 28 February 2027, on the AI Governance brief's account. Speculative Within the decade the FDA's lifecycle guidance is likely to be finalised in some form, and its real-world performance docket will either produce guidance or produce nothing; the decisive random-sample re-validation is achievable on this horizon with existing tools if anyone funds it. Speculative A second patient-endpoint screening trial in another country's screening programme is the most consequential single result available in the decade.
  • 25 yr: Speculative Either a deployment registry linking model versions to patients exists in at least one large jurisdiction, in which case failure rates per patient become computable and the authorisation count becomes interpretable, or it does not, and the list keeps measuring paperwork. Speculative Devices built on foundation models either acquire a version-pinning and revalidation convention or their validation remains a statement about a model that is no longer the one running.
  • 50 yr: Speculative Software that takes part in most clinical decisions is plausible on this horizon; whether its effect on patients is known depends on whether the measurement infrastructure was built in the first twenty-five years, because an outcome record cannot be reconstructed after the fact.
  • 100 / 250+ yr: Handwave A medicine in which every decision is model-mediated and continuously validated against outcomes is a coherent end state; this brief has no evidence about it, and the claim that it follows from the present trajectory works by assertion.

14 · Technology tree & dependencies

  • Depends on This subject waits on results that other briefs track for their own reasons: the statistical machinery and contamination controls of Intelligence Measurement, and the harmonised standards and evaluator-access questions of AI Governance. Nothing on the map blocks the decisive experiment, which needs devices, sites and a labelling protocol rather than a scientific advance.
  • Requires (not on this map) Five constraints, in the order of the tokens; sections 5 and 10 carry the figures. No regime examined here requires a deployed clinical model to report measured performance against patient outcomes on a schedule, the same missing rule the Precision Medicine brief names. No registry records which version of which model scored which patient, so no failure rate per patient can be computed. The adverse-event system cannot attribute a failure to a model version, and one report in three cannot say whether the algorithm was involved at all. For models built on the electronic record, outcome labels are partly written by the treatment the model prompts, and no general method yet removes that contamination. And nobody is paid to validate a purchased model locally: the buyer bears the cost, the vendor the reputational risk, and the patient the error.
  • Enables A working validation layer is what would let every other clinical AI subject state its effect in patient units rather than in benchmark or authorisation units. It also supplies a reference that is easy to state and hard to meet: a clinical model now has to show its performance before the clinical action it is meant to inform, a non-inferior patient endpoint at the scale of 105,934 randomised participants, and a person-and-model result rather than a model-alone result. No typed enabling edge is recorded, because that contribution is a standard rather than a technology.
  • Adjacent Human-AI Integration and Intelligence Amplification hold the human half of the clinician-and-model system; Precision Medicine holds the same accuracy-versus-outcome gap for companion diagnostics; AI Governance holds the statute.

15 · Common misconceptions & speculative claims

Established “The FDA has approved more than 1,600 AI devices, so clinical AI is proven.” Approved is the wrong verb for most of them. About 97 per cent of a 903-device sample were cleared by substantial equivalence, 55.9 per cent had a clinical performance study at clearance, and the list itself is compiled from words in authorisation summaries. Frontier The count measures regulatory throughput and says nothing about effect until it is joined to outcomes.

Established “A new name for the category means a new standing for it.” The FDA's page now notes that under Executive Order 14434 federal agencies are using “Super Intelligence” in place of “Artificial Intelligence”, so both terms may appear in its content. Frontier The devices on the list, and the evidence behind each one, are identical under either name; a label that moves upward while the evidence stays where it was is the pattern the flag system exists to catch.

Established “MASAI showed that AI cuts missed cancers by 12 per cent.” The interval-cancer proportion ratio was 0.88 with a 95 per cent interval of 0.65 to 1.18, and the trial was a non-inferiority trial. Established The accurate sentence is that AI-supported reading was not worse on interval cancers, detected more at equal specificity, and saved 44 per cent of reading workload. Frontier That is a strong result, and it is not the one in the headline.

Established “An AUROC of 0.87 means the model works.” The same sepsis model scored 0.87 counting every prediction to discharge and 0.47 counting only predictions made before clinicians had begun treatment. Frontier A discrimination figure without the time of prediction relative to the clinical action is not a measurement of usefulness.

Established “The model scores 94.9 per cent on medical scenarios, so it is ready for patients.” In the randomised test, people using the models identified relevant conditions in fewer than 34.5 per cent of cases, no better than people using their own sources. Frontier The model-alone score and the person-and-model score are different quantities, and only the second reaches a patient.

Frontier “A model validated externally once is validated.” Youssef and colleagues argue that external validation guarantees neither generalisability nor usefulness, and the decade-long replication of two sepsis feature sets shows an AUC falling from 0.729 to 0.525 through a coding change. Frontier Validation is a date-stamped statement about a site and a period, not a property of the model.

Frontier “A change control plan lets a model learn continuously from patients.” Every change a plan permits is described, bounded and validated in advance, within the device's intended use, in both the FDA and the Health Canada versions. Frontier A plan is pre-approved revision, not autonomous learning, and changes to the device's purpose fall outside it.

Frontier “Training on more diverse data fixes subgroup harm.” If the disparity lies in the labels, as Bernhardt, Jones and Glocker argue it may, more data with the same labels reproduces it; re-sampling, re-weighting and threshold calibration correct population and prevalence shift, not annotation shift. Frontier Diversity of the people in the data is necessary and not sufficient.

Established “The EU AI Act regulates medical AI from August 2027.” The Annex I date was moved to 2 August 2028, and a separate Commission proposal would move the device regulations to Section B of Annex I so that the AI Act's high-risk requirements reach devices only through the device regulations. Frontier As of November 2026 neither a European date nor a European mechanism for AI medical devices is settled.

Frontier “Recalls show the system catching failures.” In the recall data, 43.4 per cent of recalls came within twelve months of clearance, and among recalled devices from smaller public companies 96.9 per cent lacked clinical validation. Frontier A recall is a floor on failure: it captures what someone noticed and reported, and a model that degrades quietly or fails one subgroup may never produce one.

Frontier “Randomised trials of clinical AI are mostly positive, so the evidence base is strong.” Seventy of 86 trials reported favourable primary endpoints, but 46 of those endpoints were diagnostic performance or yield, 92 per cent of trials ran in a single country, and in the review that counted harm, only eight of 19 studies assessed it in a standardised way. Speculative A literature that is mostly positive on intermediate endpoints and mostly silent on harm is a literature that has not yet asked the question this brief asks.